Ohio cabinet X-ray compliance

A practical overview of verified and qualified Ohio requirements for facilities that operate covered non-medical cabinet X-ray systems.

Last reviewed for this pilot: September 25, 2026. Always open the linked official source before relying on a requirement for a specific facility.
Verified

Written radiation protection program

Ohio requires each handler of non-medical radiation-generating equipment to develop, implement, and maintain a written radiation protection program covering the subjects specified by rule.

OAC 3701:1-68-02 ↗
Qualified

Registration

Covered Ohio facilities generally must apply for registration before handling radiation-generating equipment. RayReg preserves the exemptions in the rule rather than treating registration as unconditional.

OAC 3701:1-38-03 ↗
Qualified

Cabinet operator training

Cabinet-system operators must receive training and demonstrate competence in the cabinet-specific topics in paragraph (H). Importantly, ordinary cabinet operators are exempt from paragraphs (C), (D), and (E), so RayReg does not automatically apply the general annual radiographic refresher interval.

OAC 3701:1-68-02(H) ↗
Verified

Recurring cabinet evaluation

Covered cabinet systems must be evaluated and results recorded every three months, not to exceed fourteen weeks, subject to the rule's lockout exception.

OAC 3701:1-68-06(D) ↗
Verified

Radiation surveys

Recorded radiation-area surveys are triggered at installation, during specified primary-beam maintenance/calibration procedures, and when visual inspection reveals an abnormal cabinet condition.

OAC 3701:1-68-06(E) ↗
Verified

Cabinet emission criterion

Ohio limits radiation emitted from a cabinet system to 4.4 microgray (0.5 milliroentgen) in one hour at any point five centimeters outside the external surface.

OAC 3701:1-68-06(A)(1) ↗
Important nuance

A deadline is not always permission to keep operating.

Ohio gives a seven-calendar-day repair deadline for specified failed cabinet safety components. RayReg treats that deadline together with the broader requirement that the radiation protection program prohibit operation when applicable Chapter 3701:1-68 obligations are not met. The repair window should not be presented as automatic authorization to run noncompliant equipment.
Cabinet-system ruleEquipment standards, evaluations, repairs, surveys, IRRP provisions.
Open ↗
General non-medical RGE requirementsRPP, training, records, equipment access, operating restrictions.
Open ↗
Registration ruleRegistration, exemptions, lifecycle changes, reporting.
Open ↗
Go deeper

Read the Ohio cabinet X-ray resource library.

Focused articles explain the facility-level questions behind registration, training, recurring evaluations, survey triggers, records, and failed safety devices.